Combination wave pulse 1, 2/50 – 8/20 μs (Hybrid-Surge pulse)
Pulse conforms to IEC/EN 61000-4-5
| Parameter | Value |
|---|---|
| Pulse voltage (open circuit): | ±200 V to 6.6 kV (in 1 V steps) |
| Pulse current (short circuit): | ±100 A to 3.3 kA |
| Impedance: | 2/12 Ω |
| Polarity: | positive / negative / alternate |
| Pulse repetition: | 10 s* up to 9’999 s (in 1 s steps) |
| Test duration: | 1 to 99’999 pulses, continuous |
| Phase synchronization: | asynchronous, synchronous 0 to 359º (in 1º steps) |
| Coupling: | ANSI / IEC / external |
Ring wave 0.5 µs /100 kHz
Pulse conforms to IEC/EN 61000-4-12 and ANSI (IEEE) C62.41
| Parameter | Value |
|---|---|
| Pulse voltage (open circuit): | ± 200 V to 6.6 kV (in 1 V steps) |
| Pulse current (short circuit): | ±16.6 to ±550 A, ±10%
±6.6 to ±220 A, ±10% ±1 to ±33 A, ±10% |
| Impedance: | 12/30 Ω |
| Polarity: | positive / negative / alternate |
| Pulse repetition: | 10 s* up to 9’999 s (in 1 s steps) |
| Test duration: | 1 to 99’999 pulses, continuous |
| Phase synchronization: | asynchronous, synchronous 0 to 359º (in 1º steps) |
| Coupling: | ANSI / IEC / external |
Burst (EFT) 5/50 ns
Pulse conforms to IEC/EN 61000-4-4
| Parameter | Value |
|---|---|
| Pulse amplitude: | ±200 V to 4.8 kV (in 1 V steps) – open circuit
±100 V to 2.4 kV (50 Ω matching system) |
| Burst frequency: | 100 Hz to 1000 kHz |
| Polarity: | positive / negative / alternate |
| Repetition time: | 10 ms to 9’999 ms |
| Burst time: | 0.01 ms to 9’999 ms, single pulse |
| Test duration: | 1 s to 9’999s, 1 min to 1600 min, endless |
| Phase synchonization: | asynchronous, synchronous 0 to 359º (in 1º steps) |
| Coupling: | internal / external |
Dips, Interrupts & Variations
Conforms to IEC/EN 61000-4-11, IEC/EN 61000-4-29
| Parameter | Value |
|---|---|
| Dips & Interrupts: | From EUT voltage input to 0 V, 0% (1) |
| Uvar with optional variac: | depending on model (VAR 3005A) |
| Uvar with step transformer: | 0, 40, 70, 80% (INA 650xA) |
| Peak inrush current capability: | > 500 A (at 230 V) |
| Switching times: | 1 to 5 μs (100 Ω load) |
| Event time (T-Event): | 20 µs to 9999 s, 0.5 to 9’999 cycles |
| Repetition time: | 10 ms to 9’999 ms, 1 to 9’999 s |
| Test duration: | 1 to 99’999 events, endless |
| Phase synchronization: | asynchronous, synchronous 0 to 359º (in 1º steps) |
Variation test (with VAR 3005A only)
conforms to IEC/EN 61000-4-11
| Parameter | Value |
|---|---|
| Uvar with optional variac: | up to 265 V (in 1 V steps) or up to 115% Uin (in 1% steps) |
| Decreasing time Td: | 1 ms to 9.999 ms, 0.5 to 9999 cycles, abrupt |
| Time at reduced voltage Ts: | 1 ms to 9.999 ms, 0.5 to 9999 cycles, |
| Increasing time Ti: | 1 ms to 9.999 ms, 0.5 to 9999 cycles, |
| Repetition time: | 1 s to 9’999 s |
| Events: | 1 to 99’999 |
Pulsed magnetic field in conjunction with 702A
conforms to IEC/EN 61000-4-9
| Parameter | Value |
|---|---|
| Field: | 100 to 1200 A/m |
| Polarity: | positive / negative / alternate |
| Repetition time: | 10 s to 9999s (in 1 s steps) |
| Impedance: | 2 Ω |
| Coil / impedance factor: | 0.01 to 100.00 |
| Test duration: | 1 to 9’999 pulses, endless |
| Phase synchronization: | asynchronous, synchronous 0 to 359º (in 1º steps) |
Power magnetic field in conjunction with MFT 30 or MFO 6501 and MFC 30 & MFC 300
conforms to IEC/EN 61000-4-8
| Field: | 1 to max. 40 A/m (in 1 A/m steps) |
|---|---|
| Frequency: | 50/60 Hz |
| Coil factor: | 0.01 to 99.99 |
| Test duration: | 1 to 9’999 pulses, continuous |